Memo-Graph S 7 Procedure in operation
Endress+Hauser 75
Group display
Select how the currently-selected group is to be displayed:
• Plot - time axis is plotted horizontally, signals use the entire display range.
Plot in zones - time axis is plotted horizontally, every signal is displayed in its own display zone,
signals do not overlap.
• Waterfall - time axis is plotted vertically, signals use the entire display range.
Waterfall in zones - time axis is plotted vertically, every signal is displayed in its own display zone,
signals do not overlap.
Bargraph - instantaneous values are displayed as a bargraph. The height of the bars changes in the
same way that the signal changes. Limit values are shown as marks. Digital inputs are displayed
as a square field (activated = field filled in, not activated = field not filled in).
• Digital display - instantaneous values are displayed digitally. Depending on the setting in the
device set-up, the zoom range, limit values and counter readings are alternately displayed.
History
Displays plotted measured values available in the internal memory as a curve sequence.
!
Note!
The history display can be recognised by the date/time field with black background, the “frozen”
time (seconds do not progress forward) and the changed key functions.
The displayed time refers to the time line (dividing line between curve display and measured value
display).
• Use the arrow keys (<<, <, >, >>) to move back or forward page-by-page or pixel-by-pixel
(measured value for measured value) in the time axis.
• Use “Selection” to display the values from another group at this time or “compress” the display
chronologically, i.e. display a larger time range.
• Use “Selection” to comment on a previous event, in which a text that refers to this moment is
later saved: set the desired time in the history display, press Selection - Comment and enter the
corresponding text / comment.
Analysis
Depending on the settings in the device set-up, the device can analyse the connected signals auto-
matically in defined cycles.
!
Note!
• If the device is used for batch production, the individual batches are also analysed in addition to
the selected cycles. At the device, the analysis of the current or the last batch can be called up.
• Several batches can be displayed and analysed with the associated PC software package.
Audit trail / Events
All important events are acquired and saved. The last 30 events can be viewed directly at the device.
All procedures that enable completion of processes and responsibilities are saved. They are auto-
matically completed with the user logged in at that particular time.
• System messages
• Power failures
• Limit value violations
• Login / Logout / Password change
• Product selection / start / stop
• Texts / Comments
• In/out procedures that are acquired via digital inputs
• Attaching and removing the rear-mounted cover
!
Note!
• After selecting an event, you can skip to the history display, in order to see how the event came
about or what happened with it afterwards.
• All events are listed chronologically in the PC software and can be analysed.