7 Procedure in operation Memo-Graph S
74 Endress+Hauser
7.2.3 Enter comment / text
The device can save pre-definable or free texts together with the automatically-produced messages.
A quality controller can, for example, use this to check and sign for the production.
!
Note!
The text entry can be carried out at any time, even when a production procedure is not running.
The text is not saved until a unique ID/password combination (“Electronic signature”) has been
entered.
• Each text saved includes the current date / time and indication of who saved the comment.
• If you want to add a comment about a procedure in the past, select “Extras - History”, scroll to
the desired time and then press “Selection - Comment”.
Selecting text
Select Text - Select Text and then a text from the selection list or enter a new text (“New text”).
Changing text
• Here, the selected text can be edited / adapted.
Allocating text to one or all channels
• If the comment is only related to one channel, select the corresponding channel under “Refe-
rence”. If it is a general comment (e.g. released by the quality controller) select “All channels”.
• The text is saved after a unique password/ID combination has been entered (“Electronic signa-
ture”).
7.2.4 Displaying selected group of channels / measuring points
By grouping channels you always have an overview of the status of these measuring points.
!
Note!
• Every group can contain max. 8 channels (analogue and / or digital inputs).
Use this option to display, for example, measuring points of defined system components together.
• As with the allocation of group names, the channels are also allocated into groups by the admi-
nistrator in the device set-up.
• If only one group is active, the “Group” key has no function.
Selecting a group
• Select the desired group by pressing “Group” and selecting it from the selection list.
7.2.5 Extras
Here, various options for displaying and analysing measured values, and other information menus
are available.
!
Note!
The selection of different display modes, analyses, etc. does not affect measured value acquisition,
storage, limit value monitoring, etc. All of these functions continue to run uninterrupted.
If the active group is displayed in a waterfall style, the history is also displayed in this style. In all
other display modes, the history is displayed as a curve in zones.