5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S
54 Endress+Hauser
Service: General sub-menu
Information for service technicians, e.g. for questions about the device/device fault
Set-up -> Service -> General
SW version Device software version. Please state this code if you have questions about the device.
Last power on Please state this code if you have questions about the device.
Last C-assertion Please state this code if you have questions about the device.
PRESET
" Caution!
Resets all parameters to the factory settings.
All previous values, settings and contents of the memory are deleted.
Display addresses Displays the address of the current position, as well as the help texts.
Test functions EPLD value display: instead of the scaled instantaneous value, the EPLD frequency value (raw value) is output.
Special measurement display: displays the Delta_agnd(4), Delta_agnd_U_I(5), U_P(6), U_I(7), U_U(8) measurements as
an EPLD frequency value with new analogue cards. (The settings for analogue channels 4/12 to 8/16 are converted,
these channels are automatically recorded into group 1/2.)
" Caution!
Only for service purposes.
CPU No. CPU number. Please state this code if you have questions about the device.
Unit run time Displays the device's total operating time. Please state this code if you have questions about the device.
LCD run time Displays the total operating time of the screen's back lighting. Please state this code if you have questions about the
device.
Simulation “Normal operation”: the device works with the true connected signals.
“Simulation”: the device works with simulated signals. The current device settings are taken into consideration here.
During this time, the true measured value display and storage are switched off. Instead, the simulated values are dis-
played / saved.
! Note!
If required, use the “Signal analysis- Reset” function, so that the values of the simulated signals do not corrupt your true
minima/maxima/quantities after switching back to normal operation.
If the previous signals are still required, save them first on the ATA flash card (see Chapter on “Procedure in operation -
Accessible functions - ATA flash”).