6 Guaranteeing the requirements of 21 CFR 11 Memo-Graph S
62 Endress+Hauser
6.2 Important settings at the device
6.2.1 Basic settings/operating modes
• ATA operating mode: Recommendation for stand-alone applications: stack memory
Recommendation for cyclic serial PC read-out: ring memory
•Encryption: On
•Cover: Checking
Note: After commissioning and qualification of the expected device functions/system validation,
cover the rear of the device with a terminal cover/rearside cover. The presence of the cover can
be automatically detected by means of the contact pin (setting see above).
When the cover is on/detected, even the administrator cannot make any more changes to the
set-up.
Exception: When the cover is on, text picklists can be used and administration carried out by
the logged-in administrator only. However, processes already saved are not altered.
Recommended: Seal the screws of the cover with lead. This provides practical, effective protec-
tion against manipulation/unauthorised access.
• ATAFlash change: If cyclic serial PC read-out is not activated, the warning on ATAFlash change
must be activated
Æ Basic settings Æ ATAFlash change Æ Acknowledge warn: yes
6.2.2 Signal settings/digital outputs
• Relay 1 in LV: Opening + divert
Note: With this setting, current passes through the coil of relay 1 in normal mode, the relay picks
up (creates contact between terminals 41 and 44). If the power fails and/or – if assigned in a limit
value situation – the current is switched off, the relay drops out and creates contact between ter-
minals 41 and 42. In this way, a power failure, for example, can be reliably reported
.