Memo-Graph S 10 Technical data
Endress+Hauser 89
Ingress protection class Front ingress protection:
for die-cast front with door: IP 54 (EN 60529, Cat. 2)
for stainless steel front without door: IP 65
Rear ingress protection: IP 20 (EN 60529, Cat. 2)
Electromagnetic
compatibility (EMC)
EN 61326,
NAMUR recommendation NE21:
- ESD (Electrostatic discharge): EN 61000-4-2, level 3 (6/8 kV)
- Electromagnetic interference fields: EN 61000-4-3:
Level 3 (10 V/m); additional deviation at 180 MHz: 0.7 °C (1.26 °F) at Pt100
- Burst (fast transients): EN 61000-4-4
Level 4 (2 kV signal cable / 4 kV power cable)
- Surge on power cable: EN 61000-4-5: 2 kV asymmetrical, 1 kV symmetrical
- Surge on signal cable: EN 61000-4-5: 1 kV via external protection unit
- HF by conduction: EN 61000-4-6: 10 V;
- NF by conduction: EN 61000-4-16: additional deviation at 20 kHz <0.3 %
- Power failures EN 61000-4-11: 20 ms
- Interference emission: EN 61326 class A (operation in industrial environment)
Normal mode noise rejection
EN 61298-3
40 dB at measuring range/10 (50/60 Hz ± 0.5 Hz),
not for resistance thermometer measurement
Common mode noise
rejection EN 61298-3
80 dB at 60 Vp (50/60 Hz ± 0.5 Hz)
10.7 Mechanical construction
Design, dimensions Panel mounting dimensions for version with IP 54 metal die-cast front with door and rear
panel or terminal cover, front-mounted ATA flash:
Figure 9: Panel mounting dimensions with metal die-cast front