Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 59
Please be aware that this number is not arbitrary, rather it must be defined dependent on the net-
work address of the TCP/IP network. The input form corresponds to syntax (e.g.
172.016.231.005). Complete the input with " accept".
5.10.4 Subnet mask allocation
The subnet mask must be input if the device is to establish connections in another sub-network.
Specify the subnet mask of the sub-network in which the device is located (e.g. 255.255.255.000).
Please observe: the IP address determines the class of the network. The result of this is a default
subnet mask (e.g. 255.255.000.000 for a Class B network).
5.10.5 Gateway allocation
Enter the IP address of the gateway here, if connections are to be established in other networks.
As the unit does not, at the moment, set up a connection itself via the Ethernet, it is not necessary
to enter a Gateway. Leave the set-up at "0.0.0.0".
!
Note!
Changes to the system parameters are not activated until the SETUP menu is closed and the settings
accepted. Only then will the device work with the new settings.
5.11 Communication in the network via PC software
After the device has been configured and connected to the PC network, a connection to one of the
PCs in the network can be established.
The following steps are required for this:
1. Install the supplied PC software on the PC via which communication is to take place.
2. A new device now has to be registered in the database.
After the device description has been input, select how the device settings are to be transferred.
In this case, select Ethernet (TCP/IP):