Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 53
Interfaces: Data interface (option) sub-menu
Selecting the data interface via which communication is made.
Settings only required if you are communicating via an external Profibus module.
5.7.8 Service
Displays and settings for calibration.
"
Caution!
Changes must only be made by qualified skilled personnel. Incorrect settings cause malfunctions.
Loss of claim to guarantee.
Set-up -> Interfaces -> Data interface
Communication with Not used: the data interface is not used.
Ext. DP slave module: communication via the externally-connected DP slave module. Connection to RS 232 interface
Slave address Slave address entry.
Baudrate Transmission rate on DP side. This is automatically detected or can be set between “45.45 kBaud” and “12 Mbaud”.
Timeout If no measured value is received from the bus within the set time, bus operation is defective (or incorrect settings). The
device displays this with a message. In this case, no measured values can be registered.
Input/Output PLC Reference data structure: “100/60”. 100 Bytes are transferred to the PLC, 60 Bytes from the PLC. If this changes, the
power supply to the Profibus coupler must be switched off and on.
See also “Analogue inputs - Signal” and “Analogue inputs -Data interface”