Promag CON-485 Series Operations Instructions Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 98
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 57
5.9.4 Commissioning a modem link
In principle, any conventional modem with AT command set can be used for data transfer between
your device with RS 232 interface and the supplied PC software.
!
Note!
Data transmission between an analogue modem and an ISDN terminal is not possible.
Modem at device
The modem, which is later connected to the device, must be initialised once with the PC software
(Miscellaneous - Preparing modem for device). The modem is, therefore, connected to the PC with
its original cable (normally enclosed with every modem).
Initialisation must be carried out with the same data format (baudrate, data bits, parity) with which
the measuring device works.
After successful initialisation, the modem is connected to the device with a special (zero) modem
cable.
Only three cables are required (TxD, RxD, GND).
Cable assignment:
Figure 6: Cable assihnment modem
!
Note!
The modem's original cable can not be used for this because the device and the modem have the
same pin assignments at the interface plug.
Modem at PC
The modem, which works from the PC, does not have to be initialised. Connection to the PC is
made with the (normally included with the modem) original modem cable.
The first connection to the receiver is made as follows:
In the PC software, select “Display/change device settings - New device”
Select device, set interface parameters manually (COM, baudrate, number of data bits, parity)
Activate modem operation - Set up modem
Enter receiver's telephone number.
–OK
Now enter the telephone number under which the device connected by modem can be reached and
start the connection with “OK”.
5.10 Set up Ethernet connection via internal interface
Before a connection can be established via the PC network, the system parameters in the device
must be set.
SUB-D nine pin device SUB-D nine pin modem
22
33
55
SUB-D nine pin device SUB-D 25-pin modem
22
33
57
Page view 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 97 98

Comments to this Manuals

No comments